The vibration levels on my meter are not as expected

It is possible that the sensitivities outlined on the accelerometer’s certificate have not been entered into the meter correctly, or have been lost. Check the values in the menu under Function > Calibration > Channel # > Calibration by Sensitivity. The value will be in mV/ms2 and should be the same per channel as to what is written on the certificate. In Svantek meters, Channel 1 is usually X, channel 2 is Y and channel 3 is Z.